Autoguider Testing
Date/Time
20230223T1810-2330
People
Shrey, Roy, (Kasra), Onkar, Connor
Spaces
401, 433C, Dome
Goals
Figure out auto-guider
Get something other than "99 0 0" to be shown
Determine how to collect guider data on CC
Show Kasra the Moon and Mars
Test Mirror Doors
Log
Monitors were either not working or flickering, see notes.
Rebooted machine
Determined flickering was due to USB lines. We removed all but LBD1 (w/ power) and Icron Extender
We couldn't find adapters for planetary camera, so re-attached the filter wheel and planetary camera on-axis
SharpCap couldn't find FinderScope (Deep-Sky Camera)
Determined the issue was with USB connections to DC Tower.
Deep-Sky connected to LDB2, then to DC 10Gb USB port (tealISH)
On-axis camera connected to remote Icron Extender Box, which is connected to local box via CAT7, which is connected to DC 20Gb USB-C port
The ZWO Filter Wheel is connected to the remote Icron Extender Box as well
This configuration works – testing via SharpCap
Opened Dome via ASCOM
Opened the Mirror Doors without issue
Be sure to have the Mirror Doors analog switch set to "Open"
This allows DFM TCS to control the mirror doors
Pointed at the Moon, pointing was off
Shrey eye-balled pointing from dome with feedback from control room
Updated Telescope Position
Then pointed at Tycho
Updated Telescope Position
Then pointed at Mars
Updated Telescope Position
Disconnected filter wheel and planetary from on-axis
The filter wheel needed to be calibrated
J/K it was fine. SharpCap knew what was up
https://www.baader-planetarium.com/en/baader-ubvri-bessel-filter-set-%E2%80%93-photometric.html
Note: The apparent color of the filter is NOT the passband of the filter
Closed Mirror Doors
Closed Dome Shutters
There are concerns over the lower shutter overdriving after closing, but still driving until upper shutter closes
Stored filter wheel in planetary box in shelf with filter wheel cable
AUTO GUIDING EXPERIMENT
Auto Guiding Experiment
Autoguider CCD is affixed on-axis
We will follow SBIG ST4 Manual: Quick Start Guide (pgs 4-7; https://drive.google.com/file/d/1eZHGE03GMtdPJj9Op_cJs9yjb29Hh9o2/view?usp=share_link)
After plugging in Guider CCD to ST4, it showed "99 0 0"
Cardboard box is covering light baffle
Connor and Shrey are trying to find a set of settings that gives ANY other value for the readout, other than "99 0 0"
EA (Exposure Adjust) was set to its lowest value (0.1s)
b (Boost Factor) was set to its lowest value (1)
BA (Brightness Adjustment) was set to its normal value (A), which only collects data from one pixel
As opposed to "F" which sums the pixel value along with its eight neighbors
Connor and Shrey took a Dark Frame, and then switched back to "Find and Focus".
The display then showed "0 0 0"
This makes sense since the Dark Frame is subtracted from the value
Moved on to step 5 of the start up guide.
opened shutters, mirrors, and uncovered baffle
pressed "find and focus" per guide instructions
slewed to capella, eye balling it confirmed its probably capella
values still "0, 0, 0"
restarting, NEW PLAN
removed ccd, installed on-axis camera to confirm we are pointing at capella as a test star
removed on-axis camera, installed ccd
closed dome, darkened dome, covered baffle with box
beginning step 1 of quick start guide (again)
ATTEMPT 2
step 1: ccd installed and powered, currently reading "99, 0, 0"
step 2: beginning calibration, waiting two minutes
step 3: still at "99, 0, 0"
step 4: taking dark frame, dropped to "0,0,0"
step 5: "find and focus" pressed, removed box, opening shutters, already slewed to capella
still seeing "0,0,0"
next steps?
deviating from the manual, going to try and adjust EA to the maximum value and see if we can get anything besides 0
roy raised a point, if the max value is 99 and we are using the value as the dark value, no matter what we get we could just be subtracting 99 and getting 0 for every capture we take
SUCCESS? we got a value of 99
display refreshes every 20 seconds
reduced EA to 1 second
still getting 99
reduced EA to 0.1 seconds
getting 99???
this may be because when we change certain settings, we need to take a new dark frame
changing EA to 20 seconds, shutting hatch, taking another dark frame
(0,0,0) is being displayed
opening shutters, uncovering baffle, attempting another "find and focus"
at least three frames had been taken, still displaying "0,0,0"
Attempting to take image via CCD
shrey, onkar and connor split ways looking into software to utilize the ccd
onkar:
i found this, it looks promising but i could not get it to work
in CCDOps5, there is an autoguide option under "track" in the top command bar
could be promising if we ever get the software to work
connor:
found the above, Forum from 2009 saying the software can't be used on anything after Windows 98SE
can emulate win98 via virtualbox
will be helpful for shrey's tab as well
shrey:
d:/downloads/SBIG FTP SERVER DATA DUMP/ccd/ccd.exe
found that it could only run on win98
downloaded everything off the SBIG FTP server (nice), initial passing of the contents revealed ccd.exe mentioned in the SBIG ST-4 manual used for comm with ccd
all three of us will independently investigate these softwares
Closing up
seemed like we needed to wait for dome shutters to close all the way before mirror doors would close via tcs
switches switched off
GOOBAI
ToDo
Notes
We don't know where the adapters are for the planetary camera ↔ GAM-lite
Second DC monitor is not on. Primary monitor is flickering.
There is no set of caps/coverings for filter wheel. We feel weird leaving it in a rando
There are concerns over the lower shutter overdriving after closing, but still driving until upper shutter closes
Might have located MEADE telescope cover
Weather
Data
Related content
Direct Questions/Concerns to Roy Prouty (roy.prouty@umbc.edu)